An Implanted Blood Vessel Support Device for Arteriovenous Fistulas: A Randomized Controlled Trial

Open AccessPublished:August 22, 2019DOI:https://doi.org/10.1053/j.ajkd.2019.05.023

      Rationale & Objective

      Reducing turbulent blood flow through dialysis arteriovenous fistulas (AVFs) and radial stretching of their venous wall may attenuate hyperplasia and stenosis and improve AVF outcomes in hemodialysis patients. The goal of this study was to evaluate the safety and efficacy of the VasQ implant, which intervenes on these mechanisms by physically supporting the surgical arteriovenous anastomosis.

      Study Design

      Prospective, randomized, controlled, multicenter study.

      Settings & Participants

      40 consecutive patients with kidney failure referred for creation of a brachiocephalic fistula in 4 vascular access centers in the United Kingdom and Israel.

      Interventions

      AVF surgical creation with placement of the VasQ implant (treatment) versus AVF placement without the implant (control).

      Outcomes

      Safety assessed as percentage of severe device-related adverse events was the primary outcome. Secondary outcomes were efficacy assessments including: (1) AVF maturation at 3 months, defined as cephalic vein diameter ≥ 5 mm and flow ≥ 500 mL/min; (2) functional cumulative patency, defined as successful 2-needle cannulation for two-thirds or more of all dialysis runs for 1 month in study participants receiving dialysis; (3) cephalic vein diameter and blood flow; and (4) primary and cumulative patency at 6 months.

      Results

      No severe device-related adverse events were observed. There was no significant difference in maturation at 3 months or primary patency at 6 months between treatment and control (85% vs 80% and 80% vs 66%). Significantly larger vein luminal diameters were observed in the treatment group versus controls at 3 and 6 months (8.27 ± 2.2 vs 6.69 ± 1.8 mm [P = 0.03] and 9.6 ± 2.5 vs 7.56 ± 2.7 mm [P = 0.03]). Functional patency at 6 months was significantly greater in the treatment group (100% vs 56% [P = 0.01]).

      Limitations

      Small sample size, limited power for secondary end points.

      Conclusions

      No safety signals were detected for the VasQ external support of brachiocephalic AVFs. Higher functional patency and vein luminal diameters were achieved with the device at 3 and 6 months. VasQ may safely intervene on mechanisms associated with the disturbed hemodynamic profile in the juxta-anastomotic region.

      Funding

      Funded by Laminate Medical Technologies Ltd.

      Trial Registration

      Registered at ClinicalTrials.gov with study number NCT02112669.

      Index Words

      The functional success of arteriovenous access is the backbone of hemodialysis (HD) adequacy and patient survival.
      • Daugirdas J.T.
      • Depner T.A.
      • Inrig J.
      • et al.
      KDOQI clinical practice guideline for hemodialysis adequacy: 2015 update.
      • Lacson E.
      • Brunelli S.M.
      Hemodialysis treatment time: a fresh perspective.
      Although autogenous arteriovenous fistulas (AVFs) remain, for most patients, the preferred primary vascular access due to improved patency, reduced complications, and overall cost,
      • Young E.W.
      • Dhingra R.K.
      • Hulbert-Shearon T.E.
      • et al.
      Type of vascular access and mortality in U.S. hemodialysis patients.
      they have high rates of maturation failure, stenosis, and occlusions requiring continuous surgical and endovascular maintenance to assist maturation and maintain patency.
      Although the mechanism of AVF dysfunction remains largely unclear, observational studies of “failing-to-mature” fistulas have described several possible causes. Inherent and acquired factors related to vascular dysfunction, such as female sex, diabetes, age older than 65 years, history of peripheral vascular disease, and coronary artery disease, have been linked to early fistula failure.
      • Marie Y.
      • Guy A.
      • Tullett K.
      • Krishnan H.
      • Jones R.G.
      • Inston N.G.
      Patterns of blood flow as a predictor of maturation of arteriovenous fistula for haemodialysis.
      • Sheth R.A.
      • Freed R.
      • Tavri S.
      • Huynh TTT
      • Irani Z.
      Nonmaturing fistulae: epidemiology, possible interventions, and outcomes.
      • Masengu A.
      • Hanko J.
      Patient factors and haemodialysis arteriovenous fistula outcomes.
      Moreover, anatomical and surgical factors directly affect the functional outcome. Absence of suitable vascular anatomy, inadequate surgical technique during mobilization of the vein, and lack of surgical expertise have all been identified as contributing factors in AVF failures.
      • Shenoy S.
      Surgical technique determines the outcome of the Brescia/Cimino AVF.
      • Badero O.J.
      • Salifu M.O.
      • Wasse H.
      • Work J.
      Frequency of swing-segment stenosis in referred dialysis patients with angiographically documented lesions.
      Development of stenosis during the maturation period accounts for >90% of AVF dysfunction, and failure can occur anywhere along the fistula circuit
      • Singh P.
      • Robbin M.L.
      • Lockhart M.E.
      • Allon M.
      Clinically immature arteriovenous hemodialysis fistulas: effect of US on salvage.
      due to juxta-anastomotic stenosis in 40% and proximal venous stenosis in 60% of cases.
      • Robbin M.L.
      • Chamberlain N.E.
      • Lockhart M.E.
      • et al.
      Hemodialysis arteriovenous fistula maturity: US evaluation.
      The complex AVF geometry in conjunction with a direct exposure of the vein to high-pressure pulsatile arterial flow triggers hemodynamic changes in the juxta-anastomotic region (JAR) characterized by multidirectional flow, oscillating wall shear stress (WSS), and a substantial increase in radial forces with cyclic stretching of the intima and media.
      • Mima A.
      Hemodialysis vascular access dysfunction: molecular mechanisms and treatment.
      The complex hemodynamic profile is believed to stimulate remodeling and intimal hyperplasia–related failure through a pathway that involves endothelial cell (EC) activation and the secretion of prothrombotic and vasoconstrictive factors, known to promote intimal growth through proliferation and migration of vascular smooth muscle cells and extracellular matrix deposition.
      • Kuwahara G.
      • Hashimoto T.
      • Tsuneki M.
      • et al.
      CD44 promotes inflammation and extracellular matrix production during arteriovenous fistula maturation.
      In vitro studies have evaluated EC responses to different flow rates and related shear stress patterns,
      • Qiu Y.
      • Tarbell J.M.
      Interaction between wall shear stress and circumferential strain affects endothelial cell biochemical production.
      linking unidirectional WSS and reduced circumferential strain with EC quiescence and outward vein remodeling.
      We studied the safety and efficacy of a novel biocompatible device, VasQ (Laminate Medical Technologies Ltd), designed to reduce flow disturbances at the JAR.
      • Sheth R.A.
      • Freed R.
      • Tavri S.
      • Huynh TTT
      • Irani Z.
      Nonmaturing fistulae: epidemiology, possible interventions, and outcomes.
      • Chemla E.
      • Velazquez C.C.
      • D’Abate F.
      • et al.
      Arteriovenous fistula construction with the VasQTM external support device: a pilot study.
      The device is a blood vessel external support implant that consists of 2 nickel titanium components: (1) a laser-cut brace that wraps the artery at the JAR, regulating anastomotic angulation at 40° to 50°, and creating a 1-mm radius of curvature; and (2) an external mesh braid that determines vascular diameter and gradient (cone shape) in the first 25 mm of the juxta-anastomotic vein without contact with blood flow (Fig 1A). It is hypothesized that optimization of JAR geometry could improve a unidirectional flow pattern with uniform WSS that will favorably stabilize the hemodynamic conditions in the JAR and limit progressive intimal hyperplasia associated with undesirable inward vein remodeling.
      Figure thumbnail gr1
      Figure 1(A) The VasQ device, consisting of a brace and a braid. (B) VasQ implant in its final expanded state over the anastomosis and juxta-anastomotic region of a brachiocephalic arteriovenous fistula.
      In randomized preclinical testing using a sheep model of femoral artery to femoral vein AVF, VasQ has been shown to be safe and fistulas treated with the device remained patent at 3 months of follow-up, without exceptional injury or inflammation to the vessels.
      In a first-in-human single-center single-arm study,
      • Chemla E.
      • Velazquez C.C.
      • D’Abate F.
      • et al.
      Arteriovenous fistula construction with the VasQTM external support device: a pilot study.
      20 patients had a single-size device implanted. This first study reported freedom from device-related adverse events (AEs), 79% unassisted maturation at 3 months, 79% primary patency, and 93% functional patency in patients receiving dialysis after 6 months in brachiocephalic AVFs.
      Following the first-in-human phase, final geometric and structural modifications were implemented to improve device usability, and 3 sizes of the device were configured: model 5B for a vessel diameter range of 2.5 to 4.8 mm, model 6B for 4.8 to 5.5 mm, and model 7B for 5.5 to 6.0 mm, allowing a more accurate fit for a wide range of artery and vein diameters.
      This prospective, multicenter, open-label, randomized, controlled trial was an extension of the first-in-human study. The aim of the current study was to validate the safety and efficacy in newly created brachiocephalic AVFs with VasQ in its final design, compared to standard surgical practice.

      Methods

      Study Design

      In this prospective, multicenter, open-label, randomized, controlled trial, 40 patients were enrolled at 3 sites in the United Kingdom and 1 site in Israel between September 2015 and February 2017. Patients were equally randomly assigned to the treatment (VasQ) and control (standard surgical practice) arms and were followed up for 6 months post–AVF creation (Fig 2).
      Figure thumbnail gr2
      Figure 2CONSORT (Consolidated Standards of Reporting Trials) flow chart. Abbreviation: AVF, arteriovenous fistula.
      The study was conducted in compliance with the Declaration of Helsinki and Good Clinical Practice guidelines. All participants provided written informed consent before random allocation. The study was approved by the National Health Service Research Ethics Committee (reference number 14/EE/0062) for the UK sites and the Ethics Committee of Sheba Medical Center (reference number 2331-15-SMC) for the Israeli site.

      Patient Population

      Consecutive adult patients with kidney failure who were eligible for creation of a new brachiocephalic AVF based on a surgeon’s clinical evaluation and vein mapping Doppler ultrasound were further evaluated according to study eligibility criteria. Patients were deemed eligible if they had a minimum brachial artery and cephalic vein diameter of 3 mm at the antecubital fossa, upper arm cephalic vein depth ≤ 8 mm, with no presence of stenosis in the veins of the upper arm defined as a peak systolic velocity ratio > 2 in the outflow vein or luminal stenosis > 50% and no ipsilateral central venous stenosis (eligibility criteria are available in Item S1). Cephalic vein and brachial artery quality were categorized as excellent (absence of scarring/calcification), fair (mild to moderate scarring/calcification), or poor (severe scarring/calcification) for vein/artery, respectively.

      Randomization

      Patients were randomly assigned in a 1:1 ratio to either the control or treatment arms, using sealed randomization envelopes that were provided by an independent statistician (BioStat).

      Treatment

      All study participants underwent end-to-side brachiocephalic AVF creation under local anesthesia. Patients in the treatment group were additionally treated with implantation of the VasQ device. The device was provided sterile along with a model selection tool. A transverse 4- to 6-cm incision was performed at the antecubital fossa, followed by dissection of the cephalic vein and brachial artery. Dissection and mobilization of the cephalic vein was identical in the intervention and control groups. Venous and/or arterial side branches were tied off with ligatures, avoiding the use of surgical clips that would interfere with deployment of the device. A dedicated model selection tool was used to gauge the external diameter of the artery to aid selection of the appropriate device size. The procedure continued with division of the vein, occlusion of arterial flow, and creation of a 5- to 7-mm arteriotomy. At this stage, the VasQ was placed around the vein and held in a nonexpanded state with an atraumatic bulldog clamp. A standard end-to-side vein-to-artery anastomosis with a nonabsorbable suture, for example, polypropylene 6/0, was fashioned. The arterial and venous clamps were subsequently released, and the patency of the anastomosis was confirmed. Any leaks were corrected as required. The surgeon then performed a quick interruption of the arterial inflow to facilitate the next step. With the vein collapsed, the VasQ was deployed to its final expanded state over the vein and around the anastomosis. The device was secured in position around the artery with a nonabsorbable suture through 2 small eyelets at each edge of the arterial end of the device (Fig 1B). The arterial inflow was then restored and the AVF was inspected and palpated to confirm patency. The surgical incision was closed in routine fashion.
      Patients in the control group underwent a standard end-to-side brachiocephalic AVF procedure as described, without use of the VasQ device.

      Patient Follow-up

      On-site follow-up visits were conducted 1, 3, and 6 months postsurgery and included clinical assessment of patency and maturation, Doppler ultrasound, and review of patient records. Clinical examination included AVF inspection, palpation for the presence of thrill and vein size, auscultation for the presence of audible bruit, assessment of hand circulation, and detection of AVF-related complications. Doppler ultrasound included measurement of vein diameter, volume flow rate in the outflow vein, and detection of the presence and location of stenoses. AEs and serious AEs were recorded at each visit.

      Outcome Measures

      The primary end point of the study was freedom from severe (ie, performance of daily activities is compromised) or unanticipated device-related AEs 6 months postprocedure. Secondary efficacy outcome measures included percentage of AVFs physiologically matured at 3 months, primary and secondary patency 6 months postprocedure, and cephalic vein outflow at 1, 3, and 6 months.
      Successful physiologic maturation was defined as a minimum of a 5-mm cephalic vein diameter with volume outflow > 500 mL/min confirmed using color duplex spectral analysis.
      Primary patency was defined as percentage of patent AVF free from surgical or endovascular interventions. Secondary patency was defined as percentage of patent AVF following surgical or endovascular interventions. Patency was confirmed by the presence of an audible bruit in the cephalic vein.
      Additional analyses were performed following study completion, including comparative analysis of cephalic vein diameter (by Doppler ultrasound), cumulative functional patency at 3 and 6 months defined as successful 2-needle cannulation of patent fistulas for two-thirds or more of all dialysis runs for 1 month, and occurrence of stenosis > 50% or complete occlusion anywhere along the access vein.

      Statistical Analysis

      This was an extension phase of a first-in-human trial and so was not powered for statistical significance. Rather, the aim of the study was to provide preliminary assessment of both the safety and effectiveness of the device compared to standard surgical practice.
      Baseline characteristics and outcomes were compared between the 2 study groups. Categorical and continuous variables were compared using χ2 and 2-sided t test, respectively. Fisher exact test was used for calculating P values for binary outcomes. Continuous data were expressed as mean ± standard deviation or median and range for non-normal distributions. The level of statistical significance was set at 5%. All analyses were conducted using JMP 13.2.0 software (SAS Institute).

      Results

      Study Participants

      Overall, 40 patients were enrolled in the study, 20 in the control and 20 in the treatment group, between September 2015 and February 2017. The last patient completed 6 months’ follow-up in September 2017. Table 1 summarizes baseline characteristics by treatment group.
      Table 1Patient Demographic and Clinical Characteristics by Treatment Group
      CharacteristicTreatment (n = 20)Control (n = 20)P
      Age, y60.7 ± 12.768.9 ± 11.60.04
      Age ≥ 65 y40% (8)70% (14)0.1
      Male sex14 (70%)13 (65%)0.9
      Race0.7
       White9 (45%)10 (50%)
       Asian6 (30%)3 (15%)
       African3 (15%)5 (25%)
       Caribbean2 (10%)2 (10%)
      BMI, kg/m231.3 ± 7.229.4 ± 7.40.4
      Mean arterial blood pressure, mm Hg
       Systolic138 ± 19.5140 ± 130.7
       Diastolic71 ± 14.776 ± 12.20.2
      Cause of kidney disease0.2
       Diabetes mellitus12 (60%)9 (45%)
       Hypertensive nephrosclerosis2 (10%)1 (5%)
       Polycystic disease2 (10%)3 (15%)
       Interstitial nephritis1 (5%)0 (0%)
       Chronic pyelonephritis1 (5%)0 (0%)
       Other2 (10%)7 (35%)
      Smoking status0.3
       Never smoked9 (45%)7 (35%)
       Ex-smoker5 (25%)4 (20%)
       Current4 (20%)2 (10%)
       Unknown2 (10%)7 (35%)
      Significant comorbid conditions
       Diabetes13 (65%)12 (60%)0.7
       Hypertension19 (95%)18 (90%)0.5
       Coronary/cardiac disease5 (25%)4 (20%)0.7
       Cerebrovascular disease1 (5%)0 (0%)0.2
      Previous access in the same limb5 (25%)3 (15%)0.4
      Dialysis status
       Active dialysis at screening9 (55%)9 (55%)0.9
       CVC use at screening6 (67%)8 (89%)0.3
      Total prior access procedures (all limbs and types)0.9
       06 (30%)8 (40%)
       19 (45%)8 (40%)
       ≥25 (25%)4 (20%)
      Prior access procedures in study limb0.4
       015 (75%)17 (85%)
       ≥15 (25%)3 (15%)
      Indication for index access procedure0.7
       First time access13 (65%)14 (70%)
       Failed autogenous access7 (35%)6 (30%)
      Cephalic vein external diameter, mm4.0 ± 1.244.0 ± 1.00.7
      Brachial artery external diameter, mm4.1 ± 0.774.5 ± 1.10.2
      Cephalic vein quality0.7
       Excellent12 (60%)13 (65%)
       Fair8 (40%)7 (35%)
       Poor0 (0%)0 (0%)
      Brachial artery quality0.5
       Excellent16 (80%)16 (80%)
       Fair4 (20%)3 (15%)
       Poor0 (0%)1 (5%)
      Abbreviations: BMI, body mass index; CVC, central venous catheter.
      Mean age was significantly younger in the treatment group, although the percentage of patients 65 years or older was not. No other statistically significant differences in baseline characteristics were observed.

      Surgical Procedure

      VasQ was successfully implanted in 100% (20/20) of patients randomly assigned to the treatment arm with no intraoperative exclusions. Table 2 summarizes intraoperative characteristics of blood vessels and the implantation procedure.
      Table 2Intraoperative Characteristics of Brachiocephalic AVF Procedure
      CharacteristicTreatment (n = 20)Control (n = 20)P
      Skin incision to skin closure time, min58 ± 0.1749 ± 0.270.2
      Cephalic vein diameter, mm4.0 ± 1.244.0 ± 1.00.7
      Brachial artery diameter, mm4.1 ± 0.774.5 ± 1.10.2
      Cephalic vein quality0.7
       Excellent12 (60%)13 (65%)
       Fair8 (40%)7 (35%)
       Poor0 (0%)0 (0%)
      Brachial artery quality0.5
       Excellent16 (80%)16 (80%)
       Fair4 (20%)3 (15%)
       Poor0 (0%)1 (5%)
      Access limb
       Right8 (40%)6 (30%)0.5
       Dominant7 (35%)5 (25%)0.5
      VasQ device model
      Model 5B is for a vessel diameter range of 2.5 to 4.8mm; model 6B, for 4.8 to 5.5mm; and model 7B, for 5.5 to 6.0mm.
       5B13 (65%)NA
       6B5 (25%)NA
       7B2 (10%)NA
      Abbreviations: AVF, arteriovenous fistula; NA, not applicable.
      a Model 5B is for a vessel diameter range of 2.5 to 4.8 mm; model 6B, for 4.8 to 5.5 mm; and model 7B, for 5.5 to 6.0 mm.
      No intraoperative device–related AEs were reported. Mean total surgical time (from incision to skin closure) was not significantly different: 58 ± 17 versus 49 ± 27 minutes for treatment versus control, respectively.

      Safety

      No severe device-related AEs were reported during the study. Two control patients with a background of known cardiac disease died between the 3- and 6-month follow-up due to cardiac events (cardiac arrest and asystole), both with a functioning fistula at the time of death. No deaths occurred in the treatment group. There were no statistically significant differences between the treatment group and controls in access-related AEs (Table 3).
      Table 3Access-Related Adverse Events
      EventTreatment (n = 20)Control (n = 20)P
      All-cause mortality0 (0%)2 (10%)0.5
      AVF-related mortality0 (0%)0 (0%)0.9
      AVF revascularization2 (10%)2 (10%)0.9
       Thrombectomy1 (5%)0 (0%)
       Fistuloplasty1 (5%)2 (10%)
      Access site complications0.7
       Loss of AVF patency3 (15%)4 (20%)
       Thrombosis3 (15%)2 (10%)
       Surgical ligation0 (0%)2 (10%)
      Steal syndrome2 (10%)2 (10%)0.9
      Postoperative wound infection0 (0%)0 (0%)0.9
      Postoperative bleeding0 (0%)0 (0%)0.9
      Neuropathy5% (1)0 (0%)0.3
      Hypoesthesia1 (5%)2 (10%)0.5
      Arm swelling0 (0%)1 (5%)0.3
      Abbreviation: AVF, arteriovenous fistula.
      No episodes of access-site infection occurred in either arm. Two patients in each group developed steal syndrome. Events in the treatment arm were graded as mild and moderate and did not require intervention. In the control arm, events were graded as moderate and severe, requiring surgical ligation of 1 AVF due to hand ulceration.

      AVF Maturation

      Proportions of patients with assisted and unassisted maturation and physiologic variables of the AVFs are reported in Table 4. There was no difference in achieving successful assisted and unassisted maturation between the treatment and control arms 1 and 3 months postsurgery.
      Table 4AVF Outcomes and Physiologic Characteristics
      Outcome/CharacteristicTreatmentControlP
      Assisted maturation
      Physiologic maturation defined as patent fistula with cephalic vein diameter≥5mm and AVF volume flow rate>500mL/min.
       1 mo85% (17/20)85% (17/20)0.9
       3 mo85% (17/20)80% (16/20)0.9
      Unassisted maturation
      Physiologic maturation defined as patent fistula with cephalic vein diameter≥5mm and AVF volume flow rate>500mL/min.
       1 mo80% (16/20)80% (16/20)0.9
       3 mo80% (16/20)80% (16/20)0.9
      Functional patency
      Successful establishment on hemodialysis through study AVF using 2-needle cannulation in patients actively dialyzed with a patent fistula for two-thirds or more of all dialysis runs for 1 month.
       3 mo90% (9/10)45% (5/11)0.06
       6 mo100% (14/14)56% (5/9)0.01
      Cephalic vein volume flow, mL/min
       1 mo1,259.06 ± 398.61,208.35 ± 543.20.8
       3 mo1,500.71 ± 518.91,113.5 ± 661.60.06
       6 mo1,393.7 ± 673.61,046.88 ± 625.50.1
      Cephalic vein diameter, mm
       1 mo6.94 ± 1.46.65 ± 1.30.5
       3 mo8.27 ± 1.36.69 ± 1.80.03
       6 mo9.6 ± 2.57.56 ± 2.70.03
      AVF patency
       Primary patency at 6 mo80% (16/20)66% (12/18)0.5
       Secondary patency at 6 mo85% (17/20)77% (14/18)0.6
      Note: Values given as percentage (number affected/number at risk) or mean ± standard deviation.
      Abbreviation: AVF, arteriovenous fistula.
      a Physiologic maturation defined as patent fistula with cephalic vein diameter ≥ 5 mm and AVF volume flow rate > 500 mL/min.
      b Successful establishment on hemodialysis through study AVF using 2-needle cannulation in patients actively dialyzed with a patent fistula for two-thirds or more of all dialysis runs for 1 month.
      At 3 months postsurgery, functional patency was achieved in 90% (9/10) of treated patients established on HD versus 45% (5/11) in the control group (P = 0.06). At the end of the study (6 months’ follow-up), 100% (14/14) functional patency was achieved in the treatment group versus 56% (5/9) in the control group (P = 0.01).
      Mean cephalic vein lumen diameters, measured using duplex ultrasound ∼8 cm proximal to the anastomosis (a common cannulation site for HD), 3 and 6 months post–AVF creation were 8.27 ± 2.2 versus 6.69 ± 1.8 mm (P = 0.03) and 9.6 ± 2.5 versus 7.56 ± 2.7 mm (P = 0.03), for device versus control, respectively.
      Mean AVF volume flow rates measured at the above point at 1, 3, and 6 months in the treatment and control groups were 1,259.06 ± 398.6 versus 1,208.35 ± 543.2 mL/min (P= 0.8), 1,500.71 ± 518.9 versus 1113.5 ± 661.6 mL/min (P= 0.06), and 1,393.7 ± 673.6 versus 1046.88 ± 625.5 mL/min (P= 0.1), respectively.

      AVF Patency

      There was no statistically significant difference in primary and cumulative patency between the treatment and control arms 6 months postsurgery. Primary patency was achieved in 80% (16/20) versus 66% (12/18) in treatment and control, respectively (P = 0.5). Cumulative patency was 85% (17/20) in the treatment group versus 72% (13/18) in the control group (P = 0.6).
      Stenosis, defined as >50% reduction in luminal diameter anywhere between the anastomosis and the cephalic arch (inclusive), measured using Doppler, was detected in 15% (3/20) of patients in the treatment arm versus 50% (10/20) in the control group (P = 0.04) during the follow-up period. Although the incidence of stenosis was not associated with a significant decrease in primary and cumulative patency, mean AVF volume flow rate and vein diameter at 6 months were significantly lower in AVFs with stenosis versus AVFs free from stenosis (698.8 ± 560.2 vs 1,488.9 ± 552.3 mL/min [P = 0.001] and 6.07 ± 3.7 vs 8.92 ± 3.1 mm [P = 0.03]).

      Discussion

      We report on an initial assessment of the safety and potential of VasQ, an external support device, to improve maturation and patency outcomes by optimizing the geometrical configuration of the JAR with the aim of reducing flow disturbance. Disturbed flow patterns in the JAR are characterized by increased turbulence, oscillating WSS, and increased wall tension in the JAR of the outflow vein. Suboptimal hemodynamic patterns that develop in response to the abnormal artery-vein connection have been described as key factors associated with undesirable inward vein remodeling
      • Remuzzi A.
      • Bozzetto M.
      • Brambilla P.
      Is shear stress the key factor for AVF maturation?.
      through the development of neointimal hyperplasia resulting in stenosis and eventually occlusion of the AVF. This rapid formation of a neointimal layer occurs in response to the hemodynamic changes that promote EC activation, increased expression of growth factors, and secretion of prothrombotic and vasoconstrictive substances.
      • Mima A.
      Hemodialysis vascular access dysfunction: molecular mechanisms and treatment.
      • Lee T.
      • Roy-Chaudhury P.
      Advances and new frontiers in the pathophysiology of venous neointimal hyperplasia and dialysis access stenosis.
      The association between geometrical characteristics of the arteriovenous anastomosis and AVF outcomes has been investigated in several retrospective clinical studies and computational fluid dynamics models. However, to date, these have not been translated to a technology that improves clinical outcomes. Sadaghianloo et al
      • Sadaghianloo N.
      • Jean-Baptiste E.
      • Rajhi K.
      • et al.
      Increased reintervention in radial-cephalic arteriovenous fistulas with anastomotic angles of less than 30 degrees.
      associated anastomotic angles <30° with decreased patency and higher reintervention rate in radial-cephalic AVFs. Anastomotic angle had no effect on outcomes of brachiocephalic fistulas; however, the median angle reported for brachiocephalic fistulas was 90° (range, 10°-170°), suggesting a practical limitation in controlling the anastomotic angle. Analysis of WSS profiles in side-to-side versus end-to-side anastomosis by Hull et al
      • Hull J.E.
      • Balakin B.V.
      • Kellerman B.M.
      • Wrolstad D.K.
      Computational fluid dynamic evaluation of the side-to-side anastomosis for arteriovenous fistula.
      concluded that the most uniform WSS profile occurs in side-to-side anastomoses, followed by 45° end-to-side and then 90° end-to-side configuration.
      Similarly, an implantable anastomotic connector made of siliconized polyurethane (OptiFlow, Bioconnect Systems Inc) with a predefined angle and cross-sectional area has been tested.
      • Chemla E.
      • Tavakoli A.
      • Nikam M.
      • et al.
      Arteriovenous fistula creation using the Optiflow vascular anastamotic connector: the OPEN (Optiflow PatEncy and MaturatioN) study.
      • Nikam M.
      • Chemla E.S.
      • Evans J.
      • et al.
      Prospective controlled pilot study of arteriovenous fistula placement using the novel Optiflow device.
      Although there are some conceptual similarities between the Optiflow and the VasQ, there are several fundamental and possibly crucial differences between the 2. Use of OptiFlow dictated a change in surgical technique with the use of special instruments and intravascular placement of synthetic material; VasQ does not require the introduction of prosthetic material to the bloodstream and deployment of the device over the arteriovenous anastomosis is performed using standard surgical equipment. In terms of efficacy, OptiFlow failed to demonstrate a clinical benefit in functional patency rates with an associated high rate of outflow vein thrombosis.
      In this first randomized post–CE (Conformité Européenne) mark study, no severe or serious device-related AEs were reported. No differences were observed between treatment and control with regard to the type, severity, and rate of complications. Complete loss of patency at the end of the 6-month follow-up period was reported in 15% of treated patients and 25% of controls, with no postoperative wound infection or bleeding in any of the study groups. The safety profile reported here is consistent with reports from first-in-human, single-arm, pre–CE mark studies in which no device-related AEs or serious AEs were reported. With a small sample size, our study was not powered to identify differences in AEs. However, because the device is completely external to the blood vessels, covers a limited segment of the JAR without extending into the cannulation segment, and is made of nickel titanium, a compound with well-established biocompatibility, it is not expected to increase the rate of known complications or introduce new risks.
      Patient demographics and baseline medical background were well-balanced between the control and treatment groups for most variables. However, mean patient age was younger in the treatment (60.7 ± 12.7 years) compared to the control (68.9 ± 11.6 years) group (P = 0.04). Although age has been reported as an independent risk factor for AVF failure,
      • Misskey J.
      • Faulds J.
      • Sidhu R.
      • Baxter K.
      • Gagnon J.
      • Hsiang Y.
      An age-based comparison of fistula location, patency, and maturation for elderly renal failure patients.
      several studies have failed to confirm a negative correlation between age and AVF outcome.
      • Renaud C.J.
      • Pei J.H.
      • Lee E.J.C.
      • Robless P.A.
      • Vathsala A.
      Comparative outcomes of primary autogenous fistulas in elderly, multiethnic Asian hemodialysis patients.
      • Swindlehurst N.
      • Swindlehurst A.
      • Lumgair H.
      • et al.
      Vascular access for hemodialysis in the elderly.
      • Beaulieu M.C.
      • Dumaine C.S.
      • Romann A.
      • Kiaii M.
      Advanced age is not a barrier to creating a functional arteriovenous fistula: a retrospective study.
      In our study, age had no effect over any measured outcome, including primary patency (odds ratio [OR], 1.03; 95% CI, 0.97-1.09; P = 0.4), secondary patency (OR, 1.05; 95% CI, 0.98-1.12; P = 0.1), functional patency (OR, 0.97; 95% CI, 0.91-1.02; P = 0.3), vein diameter (β = −0.002; 95% CI, −0.09 to 0.08; P = 0.9; R2 = 0.0001), or volume flow rate (β = −6.5; 95% CI, −21.5 to 8.37; P = 0.4; R2 = 0.02).
      There was no difference between study arms in achieving successful assisted and unassisted physiologic maturation 1 and 3 months postsurgery. At 3 months postsurgery, functional patency was achieved in 90% (9/10) of patients established on HD in the treatment group versus 45% (5/11) in the control group (P = 0.06). At 6 months, functional patency was achieved in all patients established on HD in the treatment group (100% [14/14]) versus 56% (5/9) in the control group (P = 0.01).
      AVFs have been proven to be superior to prosthetic grafts and tunneled lines in maintenance HD patients. Current guidelines dictate that the majority of incident patients should commence dialysis through a functioning AVF.
      Vascular Access 2006 Work Group
      Clinical practice guidelines for vascular access.
      • Fluck R.
      • Kumwenda M.
      Renal Association clinical practice guideline on vascular access for haemodialysis.
      A functional AVF is an access circuit able to deliver adequate flow, usually between 350 and 400 mL/min, without recirculation for the total duration of dialysis. Several factors have been described as possible predictors of functional patency. Demographic and comorbid condition factors such as age 65 years or older, diabetes, and body mass index ≥ 27 kg/m2 were described as negative predictors,
      • Miller P.E.
      • Tolwani A.
      • Luscy C.P.
      • et al.
      Predictors of adequacy of arteriovenous fistulas in hemodialysis patients.
      as were clinical care processes such as timing of the first postoperative follow-up visit more than 2 weeks and longer from surgery to initial cannulation (OR for each additional month, 0.81; 95% CI, 0.76-0.88).
      • Allon M.
      • Imrey P.B.
      • Cheung A.K.
      • et al.
      Relationships between clinical processes and arteriovenous fistula cannulation and maturation: a multicenter prospective cohort study.
      Higher blood flow and larger vein diameter measured using ultrasound 6 weeks postsurgery were reported to be associated with successful unassisted and overall clinical maturation by Robbin et al.
      • Robbin M.L.
      • Greene T.
      • Allon M.
      • et al.
      Prediction of arteriovenous fistula clinical maturation from postoperative ultrasound measurements: findings from the Hemodialysis Fistula Maturation Study.
      Our results show no difference in outflow volume treatment and control 1, 3, and 6 months postsurgery. Mean vein lumen diameter measured at the cannulation area (8 cm proximal to the arteriovenous anastomosis) was larger in treatment vs control 3 and 6 months postsurgery. It could be postulated that the physical dimensions of the target vein are directly related to the success of cannulation; that is, larger veins are more accessible and easier to cannulate by vascular access nurses. Functional patency could also be influenced by additional factors such as cannulation skills, patient factors (eg, preference for using a central venous catheter), or study effect because this was not a blinded study. We did not detect a center effect when assessing functional patency. Of note, 3 of 4 participating centers were in the United Kingdom and contributed the vast majority of study participants (38/40). Moreover, fistula cannulation in UK dialysis centers is performed according to the clinical practice guidelines published by the British Renal Society Vascular Access Special Interest Group and Vascular Access Society of Britain & Ireland, which provide a detailed framework encouraging adherence to nursing training and sign-off, thus reducing variation among centers and limiting potential bias.
      Stenosis (>50% reduction in lumen diameter) was detected in fewer patients in the treatment arm versus control. Although this did not translate to a 50% reduction in maturation or patency, in our study, AVFs with stenosis had significantly lower flow and vein diameter compared with AVFs with no evidence of stenosis. Progression of stenotic lesions over a longer period could result in further decline in outflow volume, requiring endovascular/surgical interventions to salvage the access, causing further deterioration of the AVF. Because the device targets the JAR, where the development of stenosis is frequent, it could be postulated that reducing flow disturbances in this area could prevent the occurrence of stenosis in this area. This could assist in maintaining a more constant outflow over time and potentially reduce proximal stenosis formation as well. Confirming this hypothesis requires larger studies with longer follow-up time.
      In conclusion, in this randomized controlled study, no safety concerns were detected using VasQ external support in brachiocephalic fistulas. There was no difference between treatment and control in unassisted maturation at 3 months and primary patency at 6 months postsurgery. At 3 and 6 months, we observed improved functional patency and larger vein luminal diameter in the treatment group versus control. VasQ may prevent some of the complications associated with the disturbed hemodynamic profile in the JAR. Larger longer-term prospective studies are required to confirm the clinical benefit of the device.

      Article Information

      Authors’ Full Names and Academic Degrees

      Nikolaos Karydis, MD, PhD, FRCS, Paul Bevis, MD, FRCS, Timothy Beckitt, MD, FRCS, Daniel Silverberg, MD, Moshe Halak, MD, and Francis Calder, MD, FRCS.

      Authors’ Contributions

      Study design: NK, PB, MH; surgical procedure and data acquisition: NK, PB, TB, DS, MH, FC; data analysis: NK, PB, MH, FC. Each author contributed important intellectual content during manuscript drafting or revision and accepts accountability for the overall work by ensuring that questions pertaining to the accuracy or integrity of any portion of the work are appropriately investigated and resolved.

      Support

      Funded by Laminate Medical Technologies Ltd. The funder designed the protocol and provided data analysis. Final decision on the primary and secondary outcomes and relevant data and on the main points to be communicated in the manuscript, including the conclusions drawn, was reached in collaboration with the funder. The authors did not receive any direct payments from Laminate Medical for their activities related to this study.

      Financial Disclosure

      The authors declare that they have no other relevant financial interests.

      Data Sharing

      The data generated in this exploratory study are the property of Laminate Medical Technologies Ltd and will not be shared for the purposes of research. Personal data relevant to participating patients' future health care needs will be provided to authorized medical personnel, named by the patient and pending the patient's consent, in accordance with local rules and regulations.

      Peer Review

      Received January 3, 2019. Evaluated by 2 external peer reviewers, with direct editorial input from a Statistics/Methods Editor, an Associate Editor, and the Editor-in-Chief. Accepted in revised form May 16, 2019.

      Supplementary Material

      References

        • Daugirdas J.T.
        • Depner T.A.
        • Inrig J.
        • et al.
        KDOQI clinical practice guideline for hemodialysis adequacy: 2015 update.
        Am J Kidney Dis. 2015; 66: 884-930
        • Lacson E.
        • Brunelli S.M.
        Hemodialysis treatment time: a fresh perspective.
        Clin J Am Soc Nephrol. 2011; 6: 2522-2530
        • Young E.W.
        • Dhingra R.K.
        • Hulbert-Shearon T.E.
        • et al.
        Type of vascular access and mortality in U.S. hemodialysis patients.
        Kidney Int. 2001; 60: 1443-1451
        • Marie Y.
        • Guy A.
        • Tullett K.
        • Krishnan H.
        • Jones R.G.
        • Inston N.G.
        Patterns of blood flow as a predictor of maturation of arteriovenous fistula for haemodialysis.
        J Vasc Access. 2014; 15: 169-174
        • Sheth R.A.
        • Freed R.
        • Tavri S.
        • Huynh TTT
        • Irani Z.
        Nonmaturing fistulae: epidemiology, possible interventions, and outcomes.
        Tech Vasc Interv Radiol. 2017; 20: 31-37
        • Masengu A.
        • Hanko J.
        Patient factors and haemodialysis arteriovenous fistula outcomes.
        J Vasc Access. 2017; 18: S19-S23
        • Shenoy S.
        Surgical technique determines the outcome of the Brescia/Cimino AVF.
        J Vasc Access. 2017; 18: S1-S4
        • Badero O.J.
        • Salifu M.O.
        • Wasse H.
        • Work J.
        Frequency of swing-segment stenosis in referred dialysis patients with angiographically documented lesions.
        Am J Kidney Dis. 2008; 51: 93-98
        • Singh P.
        • Robbin M.L.
        • Lockhart M.E.
        • Allon M.
        Clinically immature arteriovenous hemodialysis fistulas: effect of US on salvage.
        Radiology. 2008; 246: 299-305
        • Robbin M.L.
        • Chamberlain N.E.
        • Lockhart M.E.
        • et al.
        Hemodialysis arteriovenous fistula maturity: US evaluation.
        Radiology. 2002; 225: 59-64
        • Mima A.
        Hemodialysis vascular access dysfunction: molecular mechanisms and treatment.
        Ther Apher Dial. 2012; 16: 321-327
        • Kuwahara G.
        • Hashimoto T.
        • Tsuneki M.
        • et al.
        CD44 promotes inflammation and extracellular matrix production during arteriovenous fistula maturation.
        Arterioscler Thromb Vasc Biol. 2017; 37: 1147-1156
        • Qiu Y.
        • Tarbell J.M.
        Interaction between wall shear stress and circumferential strain affects endothelial cell biochemical production.
        J Vasc Res. 2000; 37: 147-157
        • Chemla E.
        • Velazquez C.C.
        • D’Abate F.
        • et al.
        Arteriovenous fistula construction with the VasQTM external support device: a pilot study.
        J Vasc Access. 2016; 17: 243-248
        • Remuzzi A.
        • Bozzetto M.
        • Brambilla P.
        Is shear stress the key factor for AVF maturation?.
        J Vasc Access. 2017; 18: S10-S14
        • Lee T.
        • Roy-Chaudhury P.
        Advances and new frontiers in the pathophysiology of venous neointimal hyperplasia and dialysis access stenosis.
        Adv Chronic Kidney Dis. 2009; 16: 329-338
        • Sadaghianloo N.
        • Jean-Baptiste E.
        • Rajhi K.
        • et al.
        Increased reintervention in radial-cephalic arteriovenous fistulas with anastomotic angles of less than 30 degrees.
        J Vasc Surg. 2015; 62: 1583-1589
        • Hull J.E.
        • Balakin B.V.
        • Kellerman B.M.
        • Wrolstad D.K.
        Computational fluid dynamic evaluation of the side-to-side anastomosis for arteriovenous fistula.
        J Vasc Surg. 2013; 58: 187-194
        • Chemla E.
        • Tavakoli A.
        • Nikam M.
        • et al.
        Arteriovenous fistula creation using the Optiflow vascular anastamotic connector: the OPEN (Optiflow PatEncy and MaturatioN) study.
        J Vasc Access. 2014; 15: 38-44
        • Nikam M.
        • Chemla E.S.
        • Evans J.
        • et al.
        Prospective controlled pilot study of arteriovenous fistula placement using the novel Optiflow device.
        J Vasc Surg. 2015; 61: 1020-1025
        • Misskey J.
        • Faulds J.
        • Sidhu R.
        • Baxter K.
        • Gagnon J.
        • Hsiang Y.
        An age-based comparison of fistula location, patency, and maturation for elderly renal failure patients.
        J Vasc Surg. 2018; 67: 1491-1500
        • Renaud C.J.
        • Pei J.H.
        • Lee E.J.C.
        • Robless P.A.
        • Vathsala A.
        Comparative outcomes of primary autogenous fistulas in elderly, multiethnic Asian hemodialysis patients.
        J Vasc Surg. 2012; 56: 433-439
        • Swindlehurst N.
        • Swindlehurst A.
        • Lumgair H.
        • et al.
        Vascular access for hemodialysis in the elderly.
        J Vasc Surg. 2011; 53: 1039-1043
        • Beaulieu M.C.
        • Dumaine C.S.
        • Romann A.
        • Kiaii M.
        Advanced age is not a barrier to creating a functional arteriovenous fistula: a retrospective study.
        J Vasc Access. 2017; 18: 307-312
        • Vascular Access 2006 Work Group
        Clinical practice guidelines for vascular access.
        Am J Kidney Dis. 2006; 48: S176-S247
        • Fluck R.
        • Kumwenda M.
        Renal Association clinical practice guideline on vascular access for haemodialysis.
        Nephron Clin Pract. 2011; 118: c225-c240
        • Miller P.E.
        • Tolwani A.
        • Luscy C.P.
        • et al.
        Predictors of adequacy of arteriovenous fistulas in hemodialysis patients.
        Kidney Int. 1999; 56: 275-280
        • Allon M.
        • Imrey P.B.
        • Cheung A.K.
        • et al.
        Relationships between clinical processes and arteriovenous fistula cannulation and maturation: a multicenter prospective cohort study.
        Am J Kidney Dis. 2018; 71: 677-689
        • Robbin M.L.
        • Greene T.
        • Allon M.
        • et al.
        Prediction of arteriovenous fistula clinical maturation from postoperative ultrasound measurements: findings from the Hemodialysis Fistula Maturation Study.
        J Am Soc Nephrol. 2018; 29: 2735-2744